Troubleshooting

Error no.
Display
E001
EEPROM memory error
Instrument not calibrated,
calibration data invalid, no user
data or user data invalid
(EEPROM error), software does
E002
not match hardware (central
module)
Download error
E003
Transmitter malfunction, software
E007
does not match transmitter
version
Sensor or sensor connection
E008
faulty
No temperature sensor connected
or temperature sensor short-
E010
circuited (temperature sensor
faulty)
E025
Limit for Airset offset exceeded
E036
Sensor calibration range
exceeded
E037
Below calibration range of sensor
E045
Calibration aborted
Calibration range of installation
E049
factor exceeded
Below calibration range of
E050
installation factor
Below measuring range of main
E055
parameter
60
9
Troubleshooting
9.1

Troubleshooting instructions

The SmarTec S CLD 132 transmitter continually monitors its own functions. If the
instrument detects a defect, the error number appears on the display. This error number
is displayed underneath the main value unit display. If several errors are detected, these
can be called up with the MINUS key.
Refer to the table "System error messages" for error numbers and the appropriate
corrective measures.
In case of malfunctions or errors without error messages, use the tables "Process errors
without messages" and "Diagnosis" to locate and eliminate the error. The "Diagnosis"
table also specifies required spare parts.
9.2

System error messages

The system error messages can be called up and selected with the MINUS key.
Tests and / or measures
1. Switch instrument off and back on.
2. Set to factory settings (S11).
3. Load hardware-compatible software (with
Optoscope, see chapter "Service
equipment Optoscope").
4. If problem persists, return instrument to
your local Endress+Hauser sales agency
for repair or replace instrument.
Download must not access locked functions
(e.g. temperature table in basic version).
Check sensor and sensor connection (see
chapter "Instrument check by medium
simulation" or call the E+H Service).
Check temperature sensor and connections;
if necessary, check instrument with
temperature simulator.
Repeat Airset (in air) or replace sensor. Clean
and dry sensor before Airset.
Clean and recalibrate sensor; if necessary,
check sensor and connections.
Recalibrate.
Check pipe diameter, clean sensor and
repeat calibration.
Immerse sensor in conductive medium or
perform Airset.
SmarTec S CLD 132
Alarm contact
Error current
Factory
User
Factory
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
Endress+Hauser