System Error Messages - Endress+Hauser Levelflex M FMP40 Operating Instructions Manual

Guided level-radar
Hide thumbs Also See for Levelflex M FMP40:
Table of Contents

Advertisement

Levelflex M FMP40 FOUNDATION Fieldbus
Endress+Hauser
9.2

System error messages

Code
Description
A102
checksum error
general reset & new calibr.
required
W103
initialising - please wait
A106
downloading please wait
A110
checksum error
general reset & new calibr.
required.
A111
electronics defect
A113
electronics defect
A114
electronics defect
A115
electronics defect
A116
download error
repeat download
A121
electronics defect
W153
initialising - please wait
A160
checksum error
general reset & new calibr.
required.
A164
electronics defect
A171
electronics defect
A221
Probe pulse deviation from
average values
A241
Broken probe
Possible cause
device has been powered off before data
could be stored;
emc problem;
EEPROM defect
EEPROM storage not yet finished
processing data download
device has been powered off before data
could be stored;
emc problem;
EEPROM defect
RAM defective
ROM defective
EEPROM defective
general hardware problem
checksum of stored data not correct
no factory calibration existant; EEPROM
defective
initialisation of electronics
device has been powered off before data
could be stored;
emc problem;
EEPROM defect
hardware problem
hardware problem
HF module or cable between HF module
and electronics defective
Broken probe orvalue for probe length is
too long
probe break monitoring enabled without
mapping beforehand
Trouble-shooting
Remedy
reset
avoid emc problem;
if alarm prevails after reset,
exchange electronics
wait some seconds; if warning
prevails, exchange electronics
wait until warning disappears
reset
avoid emc problem;
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
Reset
if alarm prevails after reset,
exchange electronics
restart download of data
contact service
wait some seconds; if warning
prevails, power off device and
power on again
reset
avoid emc problem;
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
Check contacts on HF module
If fault cannot be eliminated:
Replace HF module
Check the probe length in 033,
Check the probe itself,if the probe is
broken, change the probe, or
change to a non contact system
disable probe break monitoring,
perform mapping and then
reactivate probe break monitoring
91