System error messages

Code
Description
A102
checksum error
total reset & new calibr.
required.
W103
initializing - please wait
A106
downloading - please wait
A110
checksum error
total reset & new calibr.
required.
A111
electronics defective
A113
electronics defective
A114
electronics defective
A115
electronics defective
A116
download error
repeat download
A121
electronics defective
W153
initializing - please wait
A160
checksum error
total reset & new calibr.
required.
A164
electronics defective
A171
electronics defective
A221
probe pulse deviates from
normal values
A241
Broken probe
Possible cause
device was switched off before data were
stored;
EMC problem;
EEPROM defective
EEPROM storage not yet finished
downloading data
device was switched off before data were
stored;
EMC problem;
EEPROM defective
RAM defective
ROM defective
EEPROM defective
general hardware problem
checksum of stored data not correct
no factory calibration available; EEPROM
cleared
initialization of electronics
device was switched off before data were
stored;
EMC problem;
EEPROM defective
hardware problem
hardware problem
HF module or cable between HF module
and electronics defective
rod probe broken, rope probe broken/torn
or value entered for probe length is too
long
probe break monitoring enabled without
mapping beforehand
Troubleshooting
Remedy
reset
avoid EMC problems;
if alarm prevails after reset,
exchange electronics
wait a few seconds;
if warning prevails, exchange
electronics
wait, message disappears after
downloading operation
reset
avoid EMC problems;
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
restart download of data
contact service
wait a few seconds; if warning
prevails, switch power off and on
again
reset
avoid EMC problems;
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
reset
if alarm prevails after reset,
exchange electronics
check contacts on HF module
if fault cannot be eliminated:
replace HF module
check the probe length in 033,
check the probe mechanically, if the
probe is broken, change the probe
or change to a non contact system
disable probe break monitoring,
perform mapping and then
reactivate probe break monitoring