Soft Windows
PW
Table 22 describes how the Soft Windows and their corresponding knobs on the
keyboard operate in
PW
PW Scanning Soft Windows Options
TABLE 22.
Window
PW Steer
(Linear transducer)
Changes the angle of
the
line within the
PW
image
2D
Angle
(Angle Correction)
Adjusts the flow direc-
tion angle
ACUSON Cypress™ Operator's Manual
scanning.
Function/Description
There are three positions: 0°, +20°, –20°. Press the
corresponding Soft Window knob to select
.
PW STEER
To angle to the right, turn the knob clockwise.
To angle to the left, turn the knob counterclock-
wise.
: When using the 7L3 transducer and both
NOTE
Color and PW are activated, only "
available in the multifunctional Soft Window. Steer-
ing
will also steer the Color Box. When
PW
Color are both active, the multifunctional Soft Win-
dow will toggle between "
Press the corresponding Soft Window knob to
select and display the angle correction cursor.
Turn the knob to adjust the incident angle to the
direction of flow. The flow direction cursor rotates
relative to the sample volume line and cursor.
The Velocity scale is automatically adjusted for
the angle displayed.
The selected angle is displayed on the screen.
Angle Correction may be activated for all
NOTE:
transducers by selecting Angle Correction on the
Setup menu and putting a checkmark in the
Enabled for all transducers box.
Pulsed Wave Doppler (PW) Scanning
Steer" will be
PW
Steer" and "
2D
C
and
2D
Steer".