DICOM Setup For Modality Worklist

If the optional networking package is installed, the
Server information fields must be completely defined before the ability to make
Modality Worklist Queries is enabled.
To specify your Modality Worklist settings, follow these steps:
Press the
function key.
SETUP
1.
Double-click the
DICOM
2.
detailed in steps 3-9:
Click the Modality Worklist Server subtitle.
3.
Click the [
] button.
ADD
4.
Identify your server by typing a user-defined name in the Server Name
5.
Field.
Set the Application Entity. This defines the server application that receives
6.
Modality Worklist queries from the Cypress system.
Set the IP Address. This is the IP address of the Modality Worklist Server
7.
where the query will be made.
Set the Port Number. This is the remote port on the server that is to receive
8.
the connection request. The default setting is 104.
The Test Result field lets you know the test result of the connection to the
9.
Modality Worklist server. This field is used in conjunction with the [
[
] buttons. Before using these buttons, verify with your IT department
VERIFY
that your Modality Worklist server will respond to a ping and a
Some systems will not respond to a ping due to network security issues.
Some Modality Worklist servers may not respond to a
mand — consult the server's
or setting up the
DICOM
[
] button. The system will test the connection and indicate its results in
PING
this field. If the ping is successful, test the
[
] button.
VERIFY
ACUSON Cypress™ Operator's Manual
Exporting Images from the Cypress System
directory listing and complete the Server Information
Conformance Statement. After changing
DICOM
server settings, test the connection by clicking the
DICOM
Modality Worklist
DICOM
PING
DICOM
Verify com-
DICOM
connection by clicking the
] and
Verify.
9-11