Siemens ACUSON Cypress Operator's Manual page 139

Table of Contents

Advertisement

Table 23 describes the menu options for the
mode.
PW Scanning Tools Functions
TABLE 23.
Button
Description
Image
Click [
IMAGE
Spectral Compression Selects Low, Medium, or High.
Low increases the Dynamic Range and displays more
shades of gray; High decreases the Dynamic Range and dis-
plays fewer shades of gray; Medium falls in between.
Transmit Power Controls the maximum acoustic output of
the transducer. Settings range from 5% to 100% of maximum.
Use the lowest setting consistent with diagnostic quality. The
current mechanical index (
acoustic output setting are displayed with the System Data.
NOTE: Refer to the "Safety, Standards, and Accuracy" section of this
manual, and the AIUM document provided with your system "Medi-
cal Ultrasound Safety" for important information about Acoustic Out-
put.
Colorize Applies a variety of colorized maps overlaying the
grayshade image. This may be useful in certain situations
PW
to optimize contrast resolution.
Text
Adds text labels to an image. See "Adding Text to an Image" on
page 5-15 for details.
Timer
Procedure Timer — A timer is available for all exams. To
access the timer, follow these steps:
Press
1.
Select [
2.
Start: Starts the timer clock.
Restart: Starts the timer clock once it has been stopped.
Stop: Stops the timer clock.
Off: Stops the timer clock and removes it from the screen.
Cancel: Closes the menu without changing the timer
clock.
ACUSON Cypress™ Operator's Manual
TOOLS
] to open a menu with these options:
MI
and click the [
TOOLS
] to display the timer.
START
Pulsed Wave Doppler (PW) Scanning
buttons in
PW
) and thermal index (
] button.
TIMER
scanning
) of the
TI
5-49