Advertisement

A r n o l d
ARCADIS Avantic
Startup
System
Start-up Instructions
Print No.:
SPR2-330.815.01.03.02
Replaces:
SPR2-330.815.01.02.02
10048590
10047680
SP
2006
© Siemens AG
The reproduction, transmission or use
of this document or its contents is not
permitted
without
express
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 12.06
written

Advertisement

Table of Contents
loading

Summary of Contents for Siemens ARCADIS Avantic

  • Page 1 A r n o l d ARCADIS Avantic Startup System Start-up Instructions 10047680 10048590 2006 © Siemens AG The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages.
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Completing protocols ..........46 Page 3 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 4 Leakage current measurement ......... 49 Completing the "ARCADIS Avantic Installation" protocol..... . 51 5 _______ Protocols______________________________________________________ 52 Protective Conductor Resistance .
  • Page 5: Notes And Symbols

    Indicates a risk of danger that if disregarded leads or may lead to a potential situation which may result in an undesirable result or state (not death, physical injury or property damage). Page 5 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 6 Should be understood as a tip. The user does not absolutely have to observe these instructions. However, there will be advan- tages if he does. Page 6 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP...
  • Page 7: Additionally Required Documents

    General Information Additionally Required Documents • ARCADIS Avantic Operator Manual • ARCADIS Avantic Wiring Diagram • IQ test (quality assurance) • Local printer (optional) assembly and setting instructions • Image intensifier laser light localizer (optional) assembly and setting instructions Page 7 of 66 Siemens AG SPR2-330.815.01.03.02...
  • Page 8: Safety Information

    If work steps must be performed using electrical power, the general safety information according to TD00-000.860.01... must be observed. Page 8 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 9: Radiation Safety Information

    Release radiation for as short a time as possible. Checks requiring the release of radiation are identified by the radiation warning symbol shown on the left. Page 9 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 10: Systems Equipped With An Integrated I.i. Laser Light Localizer

    Observe the safety information in ARTD-002.731.03... When working with the laser light localizer, do not view the laser beams directly with optical instruments. Page 10 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 11: Information On The Protective Conductor Resistance Test

    The values, including the measuring points, must be recorded as first measured values in the protective conductor resistance protocol. The measuring procedure and the measuring device used (designation and serial num- ber) must also be documented. Page 11 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 12 Separate the page with the protocol from these instructions and file it in the "Protocols" register in the system binder or log book. Fig. 1: Measuring circuit for measuring the protective conductor resistance for units that are Page 12 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP...
  • Page 13: System Leakage Current Measurement Information

    - even if the limit value of 0.2 ohms is not exceeded. System leakage current measurement information Observe the instructions in the "Safety Rules for Installation and Repair" (ARTD-002.731.17 ...). Page 13 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 14 However, please note the minimum resolution of the system leakage current measuring device and any additional manufacturer information restricting the use of the measuring device. Page 14 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP...
  • Page 15 15)), the system must be insulated during the measurement and must not be touched. Fig. 3: Measuring circuit for direct measurement of the system leakage current in compliance Page 15 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 16 DIN VDE 0751-1/2001-10, Fig. C5 for protection class I. Pos. 1 = System Pos. 2 = Application part type B (if available) Pos. 3 = Measurement setup (integrated into measuring device) Page 16 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 17 This must also be documented. Document the measuring procedure (differential measurement or direct measurement) Page 17 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 18 (insulation damage, damage from moisture, defective interference suppressor, etc.) - even if the limit value of 2.5 mA is not exceeded. Page 18 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP...
  • Page 19: Tools, Aids (Partially For Iq Test)

    97 98 596 G5321 • Resolution test 28 71 820 RE999 • Safety tester (leakage current and ground wire testing device) 51 38 727 Y0766 e.g., Unimed 1100 Page 19 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 20: Note On The Log Book

    Note on the log book The log book is located in the monitor cart behind the keyboard. It is accessible if the cover on the back of the monitor cart is removed. Page 20 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06...
  • Page 21: Notes On Adapting The Power Plug To Local Conditions

    Avantic (refer to current and voltage values stated on the line voltage label, back of monitor cart). • The required line internal resistance is reached (refer to ARCADIS Avantic Planning Guide). • The selectivity of the fuse protection is assured - according to the relevant national standards.
  • Page 22: Note On The Power Connection

    The system delivery state is provided by the power connection values on the the line volt- age label (back of the monitor cart). Necessary adjustments to local conditions are to be performed as required according to the adjustment instructions of the ARCADIS Avantic system. The system leakage current measurement guidelines must be observed. WARNING Electrical voltage! Non-compliance can lead to severe bodily injury and even death.
  • Page 23: Power Connection

    2General Startup Measuring the local line voltage • Measure the voltage and frequency at the intended outlet for the ARCADIS Avantic sys- tem. • Compare the measured line voltage with the line voltage label on the back of the moni- tor cart.
  • Page 24 General Startup Fig. 4: Monitor cart line voltage Pos. 1 Power plug Pos. 2 T1 and T2 Page 24 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 25: Function Check Of The Emergency Stop Switches

    Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm movement and emergency stop" (C-arm movement and emergency stop / p. 60). Page 25 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 26: Functional Checks

    General Startup Functional checks Operating the ARCADIS Avantic, refer to the Operator Manual • Stand and C-arm system movements • Semi-transparent slot and iris diaphragm in fluoroscopy procedure. • Collimation of the semi-transparent slot and iris diaphragm using fluoroscopy proce- dure.
  • Page 27: Functional Test Of C-Arm Lift Movement And Parking Brakes

    • Document the function of the C-arm movements, brakes and warning signals in the test protocol, "C-arm movement and emergency stop" . Page 27 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 28: System Configuration

    Please clarify in advance with the customer whether the following system configura- tions are required. Service login: Fig. 6: Service login • Select <Options> <Customer service> <Local service>. Page 28 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 29 Specific System Settings Fig. 7: Authentication_ • Enter the 6-character password and click OK (see system folder for password). Page 29 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 30: Customer Address Configuration

    Specific System Settings Customer address configuration Fig. 8: Configuration_ • Select <Configuration>. Page 30 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 31 Specific System Settings Fig. 9: Configuration • Click <Next>. Page 31 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 32 Specific System Settings Fig. 10: Local Host • Click <Site Info>. Page 32 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 33: Main System Configuration

    The marked customer-specific data may be adapted. Do not change "Serial No.". • Click <Save>. • Click <Home>. Main system configuration • In the service software, select "Main System". Page 33 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 34 Specific System Settings Fig. 12: Main_system • Click <Main system>. Fig. 13: Mainsys start • Select ARCADIS Avantic - I.I. type and click <Next>. Page 34 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 35 - Buzzer on: Buzzer sound always during radiation. The country-specific regulations must be observed for the max. fluoroscopy time and radi- ation blockage. Factory setting: Page 35 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 36: Fluoroscopy Footswitch Configuration

    Save the changes with <Save>. • Click on <Home> button. • Perform a system shutdown. • Switch on the system and wait for the system to boot. Page 36 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 37: Country-Specific Adaptation Of The Maximum Tube Current

    A maximum tube current limit is required - as far as known - only in Denmark. • Start service program and log in as described under "Service Login". Page 37 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 38 Specific System Settings Fig. 15: Main_system • Click <Main system>. Fig. 16: Mainsys start • Select ARCADIS Avantic - I.I. type and click <Next>. Page 38 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 39 Open the <Options> - <End Session> menu and click the “Shut- down system” button. Immediately thereafter, press the Power off button on the monitor cart. The system shuts down and switches off automatically. Page 39 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 40: Configuring The Network (Optional) And Remote Service

    Open the <Options> - <End Session> menu and click the “Shut- down system” button. Immediately thereafter, press the Power off button on the monitor cart. The system shuts down and switches off automatically. Page 40 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP...
  • Page 41: Backup

    Select "Backup & Restore" in the Service Software Home Menu. • Place the backup CD (located in the monitor cart service compartment) in the CD-ROM drive. Fig. 18: Backup & Restore_Command Page 41 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 42 Fig. 19: Backup & Restore_Command_Backup • Select the respective drive under Drives. - Save to CD-ROM (standard save). Fig. 20: Backup SW - Settings02 • Select "SW-Settings02" under Packages. Page 42 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 43 - Wait until "Ready” is displayed in the footer. Exiting backup • Close the window via <Home>. • Remove the backup CD-ROM from the CD-ROM drive. Page 43 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 44 It is not possible to perform a backup of the database (patient images)! Store the patient images in an archive. Saving to CD-ROM as a long-term archive is prohibited. Page 44 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP...
  • Page 45: Paper Printer (Optional)

    Plug in the USB cable. • Print a test page after the automatic test. • Follow the instructions on the CD-ROM from the manufacturer in the event of problems. Page 45 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 46: Final Work Steps

    "Measurement point 22" of column "First measured value". Mark empty fields with "n.a." as "not applicable". • In row "(1) Meas. circuit", column "First measured value", enter the text "VDE0751-1 Image C2". Page 46 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 47 Complete the "First measured value" column of the "Protective conductor resistance" protocol. • If you want to complete the protocol with additional remarks, enter them in the "Remarks" table in the protocol addendum. Page 47 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 48 Monitor cart, one screw at the right side cover Monitor cart, one screw at the left side cover Monitor cart, one screw at the right handle Monitor cart, one screw at the left handle Page 48 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP...
  • Page 49: Leakage Current Measurement

    "Measuring device, ser. no.", column "First measured value". • In "Device leakage current" protocol, row "(1) Meas. circuit", column "First measured value", enter the text "VDE0751-1 Image C5". Page 49 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 50 Complete the "First measured value" column of the "Device leakage current" protocol. • If you want to complete the protocol with additional remarks, enter them in the "Remarks" table in the protocol addendum. Page 50 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP...
  • Page 51: Completing The "Arcadis Avantic Installation" Protocol

    Final Work Steps Completing the "ARCADIS Avantic Installation" protocol. The manufacturer of this product requires information and the legislator demands proofs that a product delivered free of defects from the factory continues to possess the required and certified quality properties on installation and start-up.
  • Page 52: Protective Conductor Resistance

    Protocols Protective Conductor Resistance 5Protocols System: ARCADIS Avantic Material no.: 10048590 Serial number: ................. Customer-specific ID number ................. Tab. 7 Protective Conductor Resistance First measured value Repeat measurement values Meas. point 1: ........Meas. point 2: ........Meas. point 3: ........
  • Page 53 Measuring device, type: Meas. device ser. no.: Meas. inst. calibrated up to: Evaluation: n.a. Date: Name: Signature (*1) Meas. circuit. See (Fig. 22 / p. 54) Page 53 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 54: Measuring Circuit

    DIN VDE 0751-1/2001-10, Fig. C2. Pos. 1 Measurement setup (measuring device) Pos. 2 System Pos. 3 Application part (if applicable) Remarks: Tab. 8 Date Comments Name Signature Page 54 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 55 Protocols Date Comments Name Signature Page 55 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 56: System Leakage Current

    Protocols System leakage current System: ARCADIS Avantic Material no.: 10048590 Serial number: ................. Customer-specific ID number ................. Tab. 9 System leakage current First measured value Repeat measurement values System leakage current (Highest measured value) [mA] Line voltage during the mea-...
  • Page 57: Measuring Circuit

    DIN VDE 0751-1/2001-10, Fig. C5 for protection class II. Pos. 1 = System Pos. 2 = Application part (if applicable) Pos. 3 = Measurement setup (integrated into measuring device) Page 57 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 58 DIN VDE 0751-1/2001-10, Fig. C6 for protection class II. Pos. 1 = System Pos. 2 = Application part (if applicable) Pos. 3 = Measurement setup (integrated into measuring device) Page 58 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 59: Remarks

    Protocols Remarks: Tab. 10 Date Comments Name Signature Page 59 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 60: C-Arm Movement And Emergency Stop

    Protocols C-arm movement and emergency stop System: ARCADIS Avantic Material no.: 10048590 Serial number: ................. Customer-specific ID number ................. Tab. 11 C-arm movement and emergency stop protocol Movement Name Date Initials Yes/No Releasing/locking the brakes Downward movement, position Downward movement, position The warning signals sound as described.
  • Page 61: Network Checklist

    Protocols Network checklist System: ARCADIS Avantic Material no.: 10048590 Serial number: ................. Customer-specific ID number ................. Network checklist (1+) For SAP order no.:_______________________ Will the system have a network connection (HIS/RIS, Camera, PACS)? Mandatory for image system with network connection! To ensure that the system is preconfigured at the factory and thus optimize installation at the customer site, we request that you complete and return this list if a network connection is available (and used).
  • Page 62 TCP/IP address * Configuration/DICOM/network nodes General node properties: Logical name * Configuration/DICOM/network nodes Application entity: AE title * Configuration/DICOM/network nodes Port number * Configuration/DICOM/network nodes Page 62 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 63 * Configuration/DICOM/network nodes Retrieve yes * Configuration/DICOM/network nodes Comment: Network checklist (3-) Network data for a DICOM camera in the network (only with the DICOM Print option) Page 63 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 64 TCP/IP address * Configuration/DICOM/Print devices General node properties: Logical name * Configuration/DICOM/Print devices Application entity: AE title * Configuration/DICOM/Print devices Port number * Configuration/DICOM/Print devices Comment: Page 64 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...
  • Page 65 The specification for the buzzer setting for the USA was changed as follows: Warning time limit: changed from "45" to "50" Blocking time limit: changed from "5" to "0" Page 65 of 66 Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic Medical Solutions 12.06 CS PS SP...
  • Page 66 Changes to Previous Version Page 66 of 66 ARCADIS Avantic SPR2-330.815.01.03.02 Siemens AG 12.06 CS PS SP Medical Solutions...

Table of Contents