KINETIC TESTING

Setting parameters

REV. B 02/03 PN 51680-1
Kinetic testing always uses the central fixation target. Gaze Tracking and Blind Spot Monitoring
are not available. The current stimulus value setting will be displayed in standard Goldmann
notation in the upper, middle portion of the Kinetic Test screen.
The default settings for the kinetic stimulus are:
• Stimulus Color:
White
• Stimulus Value:
I 2 E (20 dB)
• Stimulus Speed:
4 degrees per second
Unless the default settings are changed, Kinetic testing will begin in the peripheral field at
approximately 75˚ nasally and temporally or 55˚ in the superior and inferior fields.
Each Kinetic test can have up to ten (10) sets of parameters that define different isopters or
static point stimuli. The results of the first isopter will be marked on the Kinetic test field with
the uppercase letter "A". The second isopter will be indicated with a "B" on the display. Each
subsequent new set of parameters will be marked by the next letter of the alphabet.
1
From the Kinetic Test
screen, press CHANGE
PARAMETERS.
2
Use the drop-down menus
to make any needed
changes. See the following
pages for explanations.
Press SELECTION COMPLETE
when done. You will return to
the Kinetic Test screen.
HUMPHREY FIELD ANALYZER II - i series
11-5