Zeiss HUMPHREY 720i User Manual page 258

Field analyzer ii - i series
Table of Contents

Advertisement

11-50
REV. B 02/03 PN 51680-1
4
When the screen at right
appears, press CREATE
KINETIC TEST.
5
The KINETIC PARAMETER
SETUP screen appears.
Choose the stimulus parameters
for the first isopter to be tested.
Press SELECTION COMPLETE.
6
You have the option of
entering each test point in the
order you wish to test the points
or to have the HFA II randomize
the point presentation for each
isopter. Choose YES if you wish to
have the points randomized.
HUMPHREY FIELD ANALYZER II - i series
KINETIC TESTING