Zeiss HUMPHREY 720i User Manual page 228

Field analyzer ii - i series
Table of Contents

Advertisement

11-20
REV. B 02/03 PN 51680-1
4
Press STEP BY STEP.
You will then need to select the
eye to test. The Patient Data
screen appears. Press PROCEED
when all desired patient data has
been entered.
5
Press TEST MERIDIAN to
begin the first stimulus. The
meridian to be tested is indicated
on the screen.
The stimulus will light but not
move for a short period of time.
You will then see a diamond
representing the stimulus begin to
move across the screen.
Note: The ZOOM TO 30 / EXPAND TO 80 button may be used to change the stimulus starting
point prior to starting each stimulus.
6
When the patient presses the
response button, the location
of the response will be marked
with a capital letter
corresponding to the isopter
being tested.
The next meridian to be tested is
displayed on the screen. To start
the testing of the next meridian,
press TEST MERIDIAN again.
HUMPHREY FIELD ANALYZER II - i series
KINETIC TESTING