GENERAL OPERATION
Accessing the save/transmit
option - serial connection
REV. B 02/03 PN 51680-1
1
Start at the System Setup screen. Select the SAVE / TRANSMIT OPTION. Press the TRANSMIT
VIA button to toggle between Serial and Network connection settings. Network
connections are described on the next page.
2
The drop-down menu at
the right allows you to
choose from SAVE ONLY, SAVE
AND TRANSMIT, or TRANSMIT
ONLY.
3
After choosing SAVE AND
TRANSMIT or TRANSMIT
ONLY, press SERIAL TRANSMIT
SETTINGS. The Transmit RS-232
Setup screen will appear.
4
At the Transmit RS-232
screen select the Baud Rate,
Parity, Data Bits and Stop Bits
as required for transmission.
The following choices are available:
Baud Rate
300, 600, 1200, 2400, 4800, 9600, 19200
Parity
none
Data Bits
Stop Bits
For compatibility with other Carl Zeiss Meditec products (Ensemble, HFA 600 series instru-
ments), transmit serial data using 9600 Baud at Even parity with 7 data bits and 1 stop bit.
5
Select PROCEED to save the changes or CANCEL to restore the previous values. Press
PROCEED on the first screen to save the settings and return to the System Setup screen.
odd
even
7
8
1
2
HUMPHREY FIELD ANALYZER II -
2-15
i
series