HOW TO HANDLE
DATABASE FAILURES
REV. B 02/03 PN 51680-1
It is important to realize that the HFA II has built-in safeguards against loss or damage of
valuable data. You see the evidence of this when, performing certain file functions, you see the
following pop-up window.
Database failure may occur while saving test data, saving patient data, or performing file
functions. The HFA II is equipped to detect a database failure and attempt to recover any
potentially lost information. In the event of failure, a message appears on the screen that
includes an approximate "fix" time. You may elect to address the problem immediately or
delay the process. You may want to delay the process if you have patients scheduled for testing.
In that case, you may continue testing, though all data will be saved to floppy disk only, until
the problem is corrected. The screen below reveals your choices of FIX NOW or FLOPPY ONLY.