Printout legend

Printing sequence

REV. B 02/03 PN 51680-1
Isopter points are shown as uppercase letters. They are connected to form the isopter lines.
Points that define the borders of a scotoma map, a blind spot, or a custom scan, however, are
not connected to each other. The user may connect scotoma-map points with colored pencils
and shade them. This is often done in manual Goldmann perimetry to facilitate clinical
interpretation of test results.
The following is the list of print styles for each type of test point:
Type
Style
Isopter
Uppercase
Scotoma
Outlined
Blind Spot
Outlined
Custom Scan
Inverse
Occasionally an isopter point will appear on the printout but it will not be connected to other
isopter points. This situation occurs when a meridian has been retested. Only the most recent
point on a meridian will be used for the isopter map.
To print a Kinetic test, follow the instructions below:
1
Press the PRINT FUNCTIONS
icon from any screen.
2
Designate the desired Source
and Directory Order. Select
PROCEED.
Example
A
B
C
D
D
HUMPHREY FIELD ANALYZER II - i series
KINETIC TESTING