Transferring tests from
the HFA I to the HFA II
REV. B 02/03 PN 51680-1
Transferring results between the HFA I and HFA II is done via serial cable using the
TRANSFER TESTS function.
There are certain limitations to which tests will transfer from the HFA I to the HFA II. A name
and date of birth must be included on the original HFA I test in order to transfer it
to the HFA II. The transfer process will not warn you of improperly labeled HFA I tests or test
results which are not eligible to be transferred. The following is a list of restrictions specific to
the test type being transferred:
Threshold
Fast Threshold or Master files will not transfer
Compared, Averaged or Merged tests will not transfer
Tests originally acquired on the HFA II will not transfer
The Macula test: Only the first two threshold values at each point will be displayed and printed.
Note: The Peripheral 30/60-2 test on the HFA I is the Peripheral 60-4 on the HFA II. The test
results will transfer, but the 4 points not found in the 60-4 pattern will not be included.
Screening
All screening tests will transfer from the HFA I to HFA II except:
The C-166 Screening test
A screening test where the user selected the Central or Peripheral Reference Level
Auto Diagnostic tests
Note: The Peripheral 68 test on the HFA I is the P-60 on the HFA II. These test results will
transfer, but the 8 points not found in the P-60 pattern will not be included.
Custom
Custom Arc and Profile tests will not transfer.
Kinetic
Kinetic test results cannot be transferred from the HFA I.