Final Verification

Deleting the test

REV. B 02/03 PN 51680-1
Run the new SSA Aphakic Kinetic test. Print the Numerical Values printout. Verify the meridians
(0, 45, 90, 135, 180, 225, 270, 315,), the stimulus (IV4E) and the test speed (4 degrees/sec.)
are as shown in this example. The order of the meridians presented is not significant. If the
meridians listed are not identical, you must delete the test and create a new one.
If the SSA Aphakic Test is determined to be incorrect, it must be deleted. Follow the
instructions below.
1
From the Main Menu screen, press the SYSTEM SETUP Icon. Next press ADDITIONAL
SETUP. Press CUSTOM TEST.
2
Press DELETE KINETIC TEST.
3
Choose the test you wish to delete by pressing the appropriate button. Confirm that this is
your intention.
Note: Additional information about the HFA II and kinetic testing can be found in previous
sections of Chapter 11 of the HFA II User's Guide.