Zeiss HUMPHREY 720i User Manual page 266

Field analyzer ii - i series
Table of Contents

Advertisement

11-58
REV. B 02/03 PN 51680-1
4
When the screen at right
appears, press CREATE
KINETIC TEST.
5
The KINETIC PARAMETER
SETUP screen appears. Set
the following stimulus parameters
for the isopter to be tested:
Size: IV
Intensity: 4
Intensity Modifier: E
Speed: 4
Press SELECTION COMPLETE.
6
For the SSA test, the HFA II
must randomize the point
presentation for the isopter.
Choose YES.
HUMPHREY FIELD ANALYZER II - i series
KINETIC TESTING