Hard Drive Failure: Floppy Only Option - Zeiss HUMPHREY 720i User Manual

Field analyzer ii - i series
Table of Contents

Advertisement

DATABASE MANAGEMENT
Hard drive failure:
FLOPPY ONLY option
REV. B 02/03 PN 51680-1
3
If recovery is successful
and the FLOPPY ONLY option
had never been engaged, the
HFA II automatically completes
the original function (before
failure). If recovery is successful
and the FLOPPY ONLY option had
been engaged, the HFA II displays
a reminder message. After
reading the message, select OK.
Copy to your hard drive any test
results stored to the floppy drive
since the hard drive failure.
If recovery is unsuccessful, a
message appears. Select OK.
If failure occurred while saving
test results or patient data, the
HFA II will automatically save the
data to the floppy disk. If failure
occurred while performing file
functions, go to the "Rebuild Hard
Disk Database" option.
This option is only an interim measure, to be used in the event of a hard drive failure. It is
designed to allow you to continue testing patients, and save patient and test data to floppy disk
until you can attend to the hard drive problem. By choosing this option, the hard drive
becomes disabled. When time permits, recycle power to initiate the recovery program (see
Step 4 below) or use the REBUILD HARD DISK DATABASE option described later in this
section.
1
Read the message in the
dialog box. Choose FLOPPY
ONLY.
HUMPHREY FIELD ANALYZER II - i series
9-13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Humphrey 745iHumphrey 740iHumphrey 750i

Table of Contents