To Back Up Hard Drive Data Onto Floppy Disks; To Back Up Hard Drive Data Onto Magneto-Optical Disks; Set Of Backup Floppy Disks With Labels - Zeiss HUMPHREY 720i User Manual

Field analyzer ii - i series
Table of Contents

Advertisement

9-4
To back up hard drive
data onto floppy disks
To back up hard drive data
onto magneto-optical disks
REV. B 02/03 PN 51680-1
The following describes using floppy disks to create a backup set of disks. Always remember
to use only formatted floppy disks.
1. From the File Functions screen, select BACKUP/RESTORE.
2. Choose HARD DRIVE as the Source.
3. Choose FLOPPY as the Destination.
4. Select PROCEED.
5. Insert the first floppy and select OK.
6. Insert each additional floppy disk when instructed.
7. Label the floppy disks as shown in Figure 9.5.
8. Store the floppy disks in a safe location.
Caution: Although the HFA allows you to back up the entire contents of your hard
drive to floppy disk (BACKUP to FLOPPY), Carl Zeiss Meditec does not recommend
that you adopt this as your sole backup procedure. If, while restoring data, one
floppy disk in the series is damaged, the entire set becomes unusable.
Floppy Backup
SN 745-1018
Floppy Backup
Backup Date
SN 745-1018
Disk 2 of 4
Backup Date
Disk 1 of 4
Figure 9.2: Set of Backup Floppy Disks with Labels
1. From the File Functions screen, select BACKUP/RESTORE.
2. Choose HARD DRIVE as the Source.
3. Choose MAGNETO-OPTICAL as the Destination.
4. Select PROCEED.
5. Insert the magneto-optical disk into the drive.
6. Select OK.
7. Should the HFA II require more than one magneto-optical disk, instructions will be
indicated on the screen.
8. Label and date the magneto-optical disk.
9. Store the disk(s) in a safe location.
DATABASE MANAGEMENT
Floppy Backup
SN 745-1018
Backup Date
Disk 3 of 4
HUMPHREY FIELD ANALYZER II - i series
Floppy Backup
SN 745-1018
Backup Date
Disk 4 of 4