PROBLEM

Testing Problems

Test completed immediately after
START button pressed
Full field test stops before complete

Hard Disk Problems

Message indicating hard disk failure
Patient names exist in database
without test data

Floppy Disk Problems

Floppy disk cannot be used by HFA II
Floppy disk error is indicated

Magneto-optical Disk Problems

Disk cannot be used by HFA II
Data Transfer Problems
Transfer of files between HFA I, HFA II
and HFA II - i series via serial cable
does not work
Tests missing after data transfer
Routine Cleaning
Air Intake Filter
Floppy disk drive
REV. B 02/03 PN 51680-1
R E V . A 10 9/ 9 8 P N 47653
POSSIBLE CAUSE OR SOLUTION
• SIMULATION set to ON
- Go to Additional Setup screen, set SIMULATION to OFF
• Trial lens holder in up position for peripheral portion of test
- test peripheral field without a trial lens and put holder down (3)
• Refer to "How to Handle Hard Disk Failures" for step-by-step instructions (9)
•Test results not saved with patient information
- use CLEANUP HARD DISK DATABASE feature to remove (9)
• Disk may not be properly formatted (8)
- Use INITIALIZE DISK feature (all previous data will be erased)
• Disk may be the wrong format (8)
- Make sure you use only High Density (HD) 1.44 MB floppy disks
- Do not use Super High Density (2.88 MB) disks
• Disk may be damaged or corrupted (9)
- Use REBUILD FLOPPY DATABASE feature (9)
• Magneto-optical disk may not be properly formatted (8)
- Use INITIALIZE DISK feature (all previous data will be erased)
• Disk may be the wrong format (8)
- Make sure you use only magneto-optical disks (8)
• Disk may be damaged or corrupted (9)
- Replace the magneto-optical disk (8), (12)
• Check cable attachments (8)
• RS-232 port settings are incorrect (2, 8, HFA I Manual)
• Tests were ineligible to be transferred (8)
• Clean every 3 months or when dirty (12)
• Clean every 6 months or when dirty (12)