Creating Custom Tests; Getting To The Custom Test Screen - Zeiss HUMPHREY 720i User Manual

Field analyzer ii - i series
Table of Contents

Advertisement

10-2
CREATING
CUSTOM TESTS
Getting to the
custom test screen
REV. B 02/03 PN 51680-1
This section discusses how to create your own Custom static tests. These tests can be created
as either Screening of Threshold tests. Custom tests are always created for the right eye only.
When the left eye is tested, the HFA II will flip the test pattern to accommodate for the mirrored
physiological differences. Do not attempt to design a Custom test for the left eye.
Note: You may begin testing with either the right or left eye.
Custom tests must be created by the user. Once created, they may be saved and then accessed
for testing by selecting the SHOW TEST LIBRARY button from the Main Menu screen. Up to ten
(10) Custom tests can be stored in the Custom Tests screen at any one time.
Note: SITA Standard and SITA Fast are not available for use with Custom tests.
1
Press the SYSTEM
SETUP icon from the Main
Menu screen.
2
Choose
ADDITIONAL SETUP.
HUMPHREY FIELD ANALYZER II -
CUSTOM TESTING