Cleaning The Probe Tip - Interacoustics AA222 Instructions For Use Manual

Impedance audiometer
Hide thumbs Also See for AA222:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 169
AA222 Instructions for Use – EN

4.2 Cleaning the probe tip

StepUnscrew the probe cap and remove the probe tip.
Step 2: Thread the stiff end of the cleaning brush into one of the tubes from inside.
Pull the cleaning floss completely through the probe tip tube. Clean each of the
Notice:
Only insert the cleaning brush from inside out, this will ensure that the dirt is pushed
out of the probe instead of into the probe and as well protect the gasket from being
damaged. Never clean inside the probe holes.
Diagnostic probe
Step 3: Replace the cleaned tip.
Step 4: Assemble the probe again.
Clinical probe
Page 59

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents