4.3 Cleaning the Probe Tip

Step 1: Unscrew the probe cap and remove the probe tip.
Step 2: Thread the stif f end of the cleaning brush into one of the tubes f rom inside. Pull the cleaning
f loss completely through the probe tip tube. Clean each of the three tubes. Discard f loss af ter use.
Notice:
Only insert the cleaning brush f rom inside out, this will ensure that the dirt is pushed out of
the probe instead of into the probe and as well protect the gasket f rom being damaged.
Never clean inside the probe holes.
D-0128073-C – 2022/05
AT23 5 - Instructions for Use - US
Diagnostic probe
Step 3: Replace the cleaned tip.
Step 4: Assemble the probe again.
Clinical probe
Page 41