4.2 Cleaning the probe tip

Step1: Unscrew the probe cap and remove the probe tip.
Step 2: Thread the stiff end of the cleaning brush into one of the tubes from inside.
Pull the cleaning floss completely through the probe tip tube. Clean each of the
Notice:
Only insert the cleaning brush from inside out, this will ensure that the dirt is pushed
out of the probe instead of into the probe and as well protect the gasket from being
damaged. Never clean inside the probe holes.
Diagnostic probe
Step 3: Replace the cleaned tip.
Step 4: Assemble the probe again.
Clinical probe
Page 59