AA222 Οδηγίες χρήσης – EL
Σελίδα 79
CIR 2ccm uses ANSI S3.7-1995 IEC60318-5 coupler HA2 and RETSPL uses the Insert value from comes
from ANSI S3.6 2010 and ISO 389-2 1994.
B71 / B81 uses ANSI S3.13 or IEC60318-6 2007 mechanical coupler and RETFL come from ANSI S3.6
2010 and ISO 389-3 1994. Force 5.4N ±0.5N.