Probe Tone Speaker Calibration - Interacoustics AT235 Service Manual

Impedance audiometer
Table of Contents

Advertisement

Service Manual
AT235
4.1.2

Probe tone speaker calibration

Select Probe Tone Speaker for calibration.
Connect the probe to a 2 cc coupler with corresponding preamplifier and RMS-meter.
Follow the on-screen instructions.
NOTICE
It is essential to click 'Calculate Gain'
after each frequency level calibration
Fig. 31
.
20

Advertisement

Table of Contents
loading

Table of Contents