Specifications-Fast Update Mode - Agilent Technologies 86120C User Manual

Multi-wavelength meter
Table of Contents

Advertisement

Specifications and Regulatory Information
Specifications—FAST Update Mode
Specifications—FAST Update Mode
Each laser line is assumed to have a linewidth (including modulation side-
bands) of less than 10 GHz.
All specifications apply when the instrument is in the following modes:
• FAST update mode unless noted.
page 2-
• Configured to measure narrowband devices. Specifications do not apply
when the instrument is configured to measure broadband devices.
to "Measuring broadband devices and chirped lasers" on page 2-
Wavelength
Range
Absolute accuracy (lines separated by
Differential accuracy (characteristic)
Minimum resolvable separation
Display resolution
a. Signals of equal amplitude. For lines that are separated by less than 30 GHz, wavelength accuracy is reduced.
6-8
13.
30 GHz)
a
(characteristic)
Refer to "Measurement rate" on
1270–1650
±
±
3 ppm (
0.005 nm at 1550 nm,
±
2 ppm
20 GHz (0.16 nm at 1550 nm, 0.11 nm at 1300 nm)
0.001 nm
Refer
9.
±
0.004 nm at 1310 nm)