Agilent Technologies 86120C User Manual page 92

Multi-wavelength meter
Table of Contents

Advertisement

Programming
Making Measurements
Measure delta, drift, and signal- to- noise
To select a measurement, use one of the following STATe commands:
CALC3:DELT:POW:STAT (delta power)
CALC3:DELT:WAV:STAT (delta wavelength)
CALC3:DELT:WPOW:STAT
CALC3:DRIF:STAT
CALC3:SNR:STAT
CALC3:ASNR:STAT (signal-to-noise ratio averaging)
If you select a drift measurement, you can select one of the following
additional states:
CALC3:DRIF:DIFF:STAT
CALC3:DRIF:MAX:STAT
CALC3:DRIF:MIN:STAT
CALC3:DRIF:REF:STAT
The :CALCulate3:DRIFt:PRESet command turns off the minimum, max-
imum, difference, and reference states but leaves the drift state on.
Attempting to turn more than one state on at a time results in an
"–221 Settings Conflict" error.
The *RST and SYSTem:PRESet commands turn all calculations off.
CALCulate3:PRESet turns off any CALCulate3 calculations.
3-14
(delta power and wavelength)
(drift)
(signal-to-noise ratios)
(difference)
(maximum drift)
(minimum drift)
(drift reference values)

Advertisement

Table of Contents
loading

Table of Contents