Test 5. Amplitude Accuracy And Linearity - Agilent Technologies 86120C User Manual

Multi-wavelength meter
Table of Contents

Advertisement

Test 5. Amplitude Accuracy and Linearity

Equipment
Amplitude linearity is performed using the following devices:
• 1550 nm DFB lasers
• Optical attenuator
• Agilent 11896A polarization controller
• Optical power meter
Procedure
1 Turn on the laser and allow it to warm up.
2 Connect the laser's output to the optical attenuator's input.
3 Connect the optical attenuator's output to the polarization controller's
optical input.
4 Connect the polarization controller's optical output to the optical power
meter.
5 Configure the optical power meter for 1550 nm.
6 Adjust the optical attenuator so that the power meter measures a power
level of 0 dBm.
7 Enter the attenuator setting and the measured power level in the first
row of
"Attenuator Settings" and "Power Meter Readings" columns.
8 Enter the measured power on the following line:
Pwr: ____________
9 Change the attenuator in 1 dB steps as shown in
the attenuator settings and power measurements.
After completing this step, the first two columns of the table should be
Polarization sensitivity
To ensure measurement accuracy, minimize the movement of any fiber-optic
cables during this procedure. Moving cables causes polarization changes which
affect amplitude measurements.
Table 5- 21 on page 5-
Test 5. Amplitude Accuracy and Linearity
11. Be sure to enter these values into the
Table 5-
Performance Tests
21, and enter
5-9

Advertisement

Table of Contents
loading

Table of Contents