WLIMit:STOP:WNUMber
Sets the stopping wavenumber for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP:WNUMber {?|{<real> | MINimum | MAXimum }}
<real> is a wavenumber value that is within the following limits:
Constant
MINimum
MAXimum
Attribute
Non- sequential command
Summary
Preset State: 7.87401E5 m
*RST State: 7.87401E5 m
SCPI Compliance: instrument specific
Description
This command sets the stopping range for the wavelength limit. The
default units for the <real> parameter are inverse meters. The stop
wavenumber value must be less than or equal to the start wavenumber
value or the stop wavenumber will be clipped to the start wavenumber
and a "Data out of range" error will be generated.
Description
start wavelength limit
7874 cm
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument.
to "Always force the Agilent 86120C to wait for non-sequential commands" on
page 3-12
for more information.
-1
(1270 nm)
- 1
- 1
Programming Commands
CALCulate2 Subsystem
Refer