MEASure{:ARRay | [:SCALar]} :POWer:WAVe-
length?
Returns wavelength values.
Syntax
:POWer:WAVelength? [<expected_value>[,<resolution>]]
Description
When used with a :SCALar command, a single value is returned. The
display is placed in the single- wavelength mode, and the marker is
placed on the signal having a wavelength that is closest to the
<expected_value> parameter. Default units for <expected_value>
parameter are in meters.
When used with an :ARRay command, an array of wavelengths is
returned. The display is placed in the list- by- wavelength mode.
The <resolution> parameter sets the resolution of the measurement. It
is a unitless number whose value will be limited to either 0.01 or
0.001 (whichever is closer).
Returned values are in meters. Displayed units are nanometers. Power
units are not affected.
<expected_value>
MAXimum
Constants
MINimum
DEFault
Used With
<expected_value>
SCALar
optional
ARRay
ignored
a. Although ignored, this argument must be present if the resolution argument is specified.
CONFigure command
When this function is used with the CONFigure command, the query question
mark character "?" must not be included in the string. However, the FETCh, READ,
and MEASure command are queries and require the question mark. Refer to the
examples for this command.
The highest wavelength signal
The lowest wavelength signal
The current marker position
a
Programming Commands
Measurement Instructions
<resolution>
optional
optional
4-21