Table 5-24. General SCPI Error Messages (2 of 3)
Error Number
Description
–158
"String data not allowed"
–161
"Invalid block data"
–168
"Block data not allowed"
–170
"Expression error"
–171
"Invalid expression"
–178
"Expression data not allowed"
–200
"Execution error"
–211
"Trigger ignored"
Caused by sending the *TRG command when the instrument is already
taking a measurement or when the instrument is in continuous
measurement mode.
–213
"Init ignored"
Caused by sending an INIT:IMM, READ, or MEASure command while a
measurement is already in progress or while the instrument is in
continuous measurement mode.
–221
"Settings conflict"
Caused by trying to set the instrument to a state that is not allowed. For
example, turning on drift maximum and drift minimum state
simultaneously or turning on SNR state while drift or delta state is on.
–222
"Data out of range"
–223
"Too much data"
–224
"Illegal parameter value"
– 230
"Data corrupt or stale"
Caused by trying to query measurement data immediately after a *RST
command. For example, sending *RST; FETCh or sending *RST;
:CALC2:DATA? pow.
–232
"Data questionable"
Caused by sending a resolution value in one of the measurement
functions that is outside the instrument's range.
–273
"Illegal macro label"
Reference
Error Messages