Programming Commands
CALCulate2 Subsystem
WLIMit:STOP[:WAVelength]
Sets the stopping wavelength for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP[:WAVelength] {?|{<real> | MINimum | MAXimum }}
<real> is a wavelength value that is within the following limits:
Constant
MINimum
MAXimum
Attribute
Non- sequential command
Summary
Preset State: 1650 nm
*RST State: 1650 nm
SCPI Compliance: instrument specific
Description
This command sets the stopping range for the wavelength limit. The
default units for the <real> parameter are meters. The stop wavelength
value must be greater than or equal to the start wavelength value or
the stop wavelength will be clipped to the start wavelength and a
"Data out of range" error will be generated. Setting the start wave-
length is equivalent to setting the start frequency/wavenumber because
of the inverse relationship of frequency to wavelength.
4-42
Description
start wavelength limit
1650.0 nm
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument.
to "Always force the Agilent 86120C to wait for non-sequential commands" on
page 3-12
for more information.