Programming Commands
CALCulate2 Subsystem
WLIMit:STARt:FREQuency
Sets the start frequency for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STARt:FREQuency {?|{<real> | MINimum | MAXimum }}
<real> is a frequency value that is within the following limits:
Constant
MINimum
MAXimum
Attribute
Non- sequential command
Summary
Preset State: 181.6924 THz
*RST State: 181.6924 THz
SCPI Compliance: instrument specific
Description
This command sets the starting range for the wavelength limit in hertz.
The start frequency value must be less than or equal to the stop fre-
quency value or the start frequency will be clipped to the stop fre-
quency and a "Data out of range" error will be generated. The default
units for the <real> parameter are hertz.
4-38
Description
181.6924 THz (1650 nm)
wavelength limit stop value
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument.
to "Always force the Agilent 86120C to wait for non-sequential commands" on
page 3-12
for more information.
Refer