*TRG
The *TRG (trigger) command is identical to the group execute trigger
(GET) message or RUN command.
Syntax
*TRG
Description
This command acquires data according to the current settings. This
command cannot be issued as a query. If a measurement is already in
progress, a trigger is ignored, and an error is generated.
Example
The following example starts the data acquisition according to the cur-
rent settings.
OUTPUT 720;"*TRG"
*TST?
The *TST (test) query starts a self- test on the instrument.
Syntax
*TST?
Description
The result of the test is placed in the output queue. A zero indicates
the test passed and a non- zero value indicates the test failed. The
instrument will initiate a measurement and check for any hardware
errors after the measurement is complete.
Query Response
<integer>
Example
OUTPUT 720;"*TST?"
ENTER 720;Result
PRINT Result