Agilent Technologies 86120C User Manual page 136

Multi-wavelength meter
Table of Contents

Advertisement

Programming Commands
Common Commands
*RST
The *RST (reset) command returns the Agilent 86120C to a known
condition.
Syntax
*RST
Description
For a listing of reset conditions, refer to the following table. This com-
mand cannot be issued as a query. Since this command places the
instrument in single measurement acquisition mode, any current data
is marked as invalid and a measurement query such as :FETCh?
results in error number –230, "Data corrupt or stale". You must ini-
tiate a new sweep with :INIT:IMM before you can use the :FETCh com-
mand.
Table 4-15. Conditions Set by *RST Reset (1 of 2)
4-8
Item
Display mode
Start wavelength
Stop wavelength
Graphical display
Measurement acquisition
Wavelength calibration
Elevation correction value
Wavelength units
Amplitude units
Power offset
Peak threshold
Peak excursion
Measurement speed
Setting
single wavelength
1270 nm
1650 nm
off
single
vacuum
0 meters
nm
dBm
0 dB
10 dB
15 dB
normal