Agilent Technologies 86120C User Manual page 223

Multi-wavelength meter
Table of Contents

Advertisement

{OPERation | QUEStionable}:PTRansition
Selects bits in the event register which can be set by positive transi-
tions of the corresponding bits in the condition register.
Syntax
:STATus:OPERation:PTRansition{?| <integer>}
<integer> an integer from 0 to 65535.
Attribute
Preset State: none
Summary
*RST State: none
SCPI Compliance: standard
Description
Changes in the state of a condition register bit causes the associated
OPERation Status or QUEStionable Status event register bit to be set.
This command allows you to select a positive bit transition to trigger
an event to be recognized. A positive transition is defined to occur
whenever the selected bit changes states from a 0 to a 1. You can
enter any value from 0 to 65535.
When queried, the largest value that can be returned is 32767. This is
because the most- significant register bit cannot be set true.
Example
OUTPUT 720;":STATUS:OPER:PTRansition 16"
Programming Commands
STATus Subsystem
4-95

Advertisement

Table of Contents
loading

Table of Contents