Programming Commands
CALCulate2 Subsystem
POINts?
Queries the number of points in the data set.
Syntax
:CALCulate2:POINts?
Attribute
Preset State: unaffected
Summary
*RST State: unaffected
SCPI Compliance: instrument specific
Query Only
Description
This is the number of points that will be returned by the
CALC2:DATA? query.
Query Response
For example, if six laser lines are located:
+6
PTHReshold
Sets the peak threshold limit used by the instrument to determine
valid laser line peaks.
Syntax
:CALCulate2:PTHReshold{?| {<integer> | MINimum | MAXimum | DEFault}}
<integer> represents logarithmic units in dB. Valid range is 0 to 40.
Constant
MINimum
MAXimum
4-34
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument.
to "Always force the Agilent 86120C to wait for non-sequential commands" on
page 3-12
for more information.
Value
0 dB
40 dB
Refer