Programming Commands
STATus Subsystem
{OPERation | QUEStionable}:ENABle
Sets the enable mask for the questionable or operation event register.
Syntax
:STATus:{OPERation | QUEStionable}:ENABle{?| <value>}
<integer> an integer from 0 to 65535.
Preset State: none
Attribute
Summary
*RST State: none
SCPI Compliance: standard
Description
The enable mask selects which conditions in the event register cause
the summary bit in the status byte to be set. If a bit in the enable
mask is set true and the corresponding event occurs, the summary bit
(bit 3 for the questionable status or bit 7 for the operation status) in
the status byte will be set.
Example
OUTPUT 720;":STATUS:QUESTIONABLE:ENABLE 1024"
Query Response
When queried, the largest value that can be returned is 65535. This is
because the most- significant register bit cannot be set true.