WLIMit[:STATe]
Turns wavelength limiting on and off.
Syntax
:CALCulate2:WLIMit[:STATe]{?| {ON | OFF | 1 | 0}}
Attribute
Non- sequential command
Summary
Preset State: on
*RST State: on
SCPI Compliance: instrument specific
Description
When this function is on, the Agilent 86120C has an input range from
the WLIMit STARt to the WLIMit STOP. When this function is off, the
instrument displays peaks over the full wavelength range. The graphics
display always shows the range between WLIMit:STARt and
WLIMit:STOP, regardless of the state of this command.
Whenever the Agilent 86120C receives this command, it reprocesses
the data and performs a new peak search.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument.
to "Always force the Agilent 86120C to wait for non-sequential commands" on
page 3-12
for more information.
Programming Commands
CALCulate2 Subsystem
Refer
4-37