(Set-Up Mode) - AEMC PowerPad 8435 User Manual

3-phase power quality analyzer
Hide thumbs Also See for PowerPad 8435:
Table of Contents

Advertisement

3.2 INSTRUMENT CONFIGURATION
NOTE: All configurations can also be modified through DataView
The instrument must be configured the first time it is used. The configuration is saved in memory when the instrument is turned
OFF.
1. Press the
2. Set the display language by pressing the yellow button corresponding to the screen language icons.
3. The parameter that is ready to be configured will be highlighted in yellow. To move to a different parameter, use the ▲
and ▼ buttons.
4. Press the Enter
5. Use the ▲, ▼, ◄, and ► buttons to change a value or setting; and the
6. When finished, return to the Configuration menu by pressing the
PARAMETER
Date / Time
Display
Calculation
Methods
Electrical Hook-Up
Power Quality Analyzer PowerPad
Find Quality Products Online at:
button to configure the unit. The following screen appears:
button to select a parameter.
FUNCTION
Sets the date and time format (see § 3.2.1)
Adjusts the contrast and brightness of the display
Defines the color of the voltage and current curves (see § 3.2.2)
Determines if harmonics are used or not used in calculations of reactive quantities (power and energy) - (see § 3.2.3)
• With harmonics: Harmonics are taken into account when calculating reactive parameters
• Without harmonics: Only the fundamental part is used for the calculation of reactive parameters
Determines the type of connection to the network (see § 3.2.4)
• 1-Phase 2-Wire
• 1-Phase 3-Wire
• 2-Phase 2-Wire
• 2-Phase 3-Wire
• 2-Phase 4-Wire
• 3-Phase 3-Wire
• 3-Phase 4-Wire
• 3-Phase 5-Wire
®
III Model 8435
GlobalTestSupply
www.

(SET-UP MODE)

software.
®
Figure 3-1
button.
button to save the setting.
.com
sales@GlobalTestSupply.com
15