5.4.5 RECORDINGS

The Recordings tab shows the parameters for a recording session.
Four different configurations are available. More configurations can be saved by pressing Save to File and recalled later
by pressing Load From File.
1. Select the configuration to set up: 1, 2, 3, or 4.
2. Check the box for each parameter desired in the Data to Record section.
3. Configure the harmonic values to be recorded, if desired.
4. Choose an Aggregation Period for the recording, which sets how often the recording updates while it is running.
The Model 8435 loses its scheduled recording if it is powered off before the recording begins. If it is powered off during the
recording, a partial recording will usually still exist but with the first letter of its name changed to "@". The start and end times
requested for the recording might be adjusted by the Model 8435 to be in even multiples of the averaging period. For instance,
if an integration period of 10 minutes was requested, and the start time was 9:03, the recording might not actually begin until
9:10.