AEMC PowerPad 8435 User Manual page 96

3-phase power quality analyzer
Hide thumbs Also See for PowerPad 8435:
Table of Contents

Advertisement

Measurement
SR clamp
MR clamp
MN93 clamp
SL261 clamp (10mV/A)
MN193 clamp (100A)
Peak
Current
SL261 clamp
(100mV/A)
MN193 clamp (5A)
or 5A adapter
AmpFlex
®
MiniFlex
®
Severity of Flicker (Pst)
(1)
In 1000Vrms, CAT III, provided that the voltage between each of the terminals and the ground/earth does not exceed 1000Vrms
(2)
2-phase (opposite phases) – same note as (1)
1200 x √2 ≈ 1697; 2400 x √2 ≈ 3394; 240 x √2 ≈ 339,4; 120 x √2 ≈ 169,7; 12 x √2 ≈ 16,97; 6 x √2 ≈8,485; 6500 x √2 ≈ 9192
(3)
Measurement
Excluding
AmpFlex
®
MiniFlex
®
Active
Power
(1)
AmpFlex
®
MiniFlex
®
Excluding
AmpFlex
®
MiniFlex
®
Reactive
Power
(2)
AmpFlex
®
MiniFlex
®
Apparent Power
94
Find Quality Products Online at:
Measurement range without ratio
(with unity ratio)
Min
1A
0.2A
0.1A
0.01A
0.005A
10A
0
Measurement range without ratio
(with unity ratio)
Min
Max
5mW
7800kW
(3)
5mVAR
7800kVAR
(3)
5mVA
7800kVA
(3)
GlobalTestSupply
www.
Resolution
(with unity ratio)
Max
1A
A < 1000A
1697A
(3)
1A
A ≥ 1000A
339.4A
0.1A
(3)
0.01A
A < 100A
169.7A
(3)
0.1A
A ≥ 100A
0.001A
A < 10A
16.97A
(3)
0.01A
A ≥ 10A
8.485A
0.001A
(3)
0.1A
A < 1000A
9192A
(3)
1A
A ≥ 1000A
12
0.01
Resolution
(with unity ratio)
4 digits at most
(4)
4 digits at most
(4)
4 digits at most
(4)
Power Quality Analyzer PowerPad
.com
Accuracy
±(1% + 1A)
±(1% + 1A)
±(1% + 0.1A)
±(1% + 0.01A)
±(1% + 0.01A)
±(1.5% + 5A)
See the
corresponding table
Accuracy
±(1%)
cos F ≥ 0.8
±(1.5% + 10 ct)
0.2 ≤ cos F < 0.8
(5)
±(1%)
cos F ≥ 0.8
±(1.5% + 10 ct)
0.5 ≤ cos F < 0.8
±(1%)
sin F ≥ 0.5
±(1.5% + 10 ct)
0.2 ≤ sin F < 0.5
(5)
±(1.5%)
sin F ≥ 0.5
±(2.5% + 20 ct)
0.2 ≤ sin F < 0.5
±(1%)
(5)
®
III Model 8435