6.2.4 CURRENT PROBE ACCURACY
Sensor errors are offset by a typical correction inside the device. This typical correction, applied to the phase and
amplitude, depends on the type of sensor connected (detected automatically) and the gain in the current acquisition
channel used.
The RMS current measurement error and phase error are additional errors (which must therefore be added to the device
errors), indicated as influences on the calculations carried out by the analyzer (powers, energy, power factors, tangents,
etc.).
Sensor Type
MR193
1000A
SR193
1000A
AmpFlex
®
3000A
MiniFlex
®
1000A
MN93
200A
MN193
100A
MN193
5A
5A
Adapter
SL261
10A (100mV/A)
SL261
100A (10mV/A)
N.S. = Not Specified
Power Quality Analyzer PowerPad
Find Quality Products Online at:
(TO BE ADDED TO THE MODEL 8435'S ACCURACY)
TRMS Current
[1A; 10A]
[10A; 100A]
[100A; 800A]
[800A; 1200A]
[1A; 3A]
[3A; 10A]
[10A; 100A]
[100A; 1200A]
[10A; 100A]
[100A; 6500A]
[10A; 100A]
[100A; 3000A]
[0.5A; 2A]
[2A; 10A]
[10A; 100A]
[100A; 240A]
[100mA; 300mA]
[300mA; 1A]
[1A; 120A]
[5mA; 50mA]
[50mA; 500mA]
[500mA; 6A]
[5mA; 50mA]
[50mA; 6A]
[0A; 10A]
[0A; 40A]
[40A; 100A]
®
III Model 8435
GlobalTestSupply
www.
Accuracy on IRMS
±(1.5% + 1A)
±(3%)
±(5%)
±(0.8%)
±(0.3%)
±(0.2%)
±(3%)
±(2%)
±(3%)
±(2%)
±(3% + 1A)
±(2.5% + 1A)
±(1% + 1A)
±(0.7% + 2mA)
±(0.7%)
±(1% + 0.1mA)
±(1%)
±(0.7%)
±(1%)
±(0.5%)
±(1.5% + 50mA)
±(2% + 50mA)
±(5%)
.com
Accuracy on φ
N.S.
±(2°)
±(1.5°)
N.S.
±(1°)
±(0.5°)
±(0.3°)
±(1°)
±(0.5°)
±(1°)
±(0.5°)
N.S.
±(6°)
±(3°)
±(2°)
N.S.
±(1.5°)
±(0.7°)
±(1.7°)
±(1°)
±(1°)
±(0°)
±(1°)
±(0.5°)