5.4.6 TRANSIENTS

The Transients tab allows you to set up the criteria for capturing transients.
1. In the Threshold of Voltage Difference section of the dialog box, select from the drop-down menu one of the
following options: 4L Same thresholds for all, 3L+N Same except N has its own threshold, or L1+l2+l3+N
Separate thresholds.
2. Based on the preceding selection, fill in appropriate values for L1, L2, L3, and N. (Some of these may be grayed out
and inactive, depending on the selected connection type.)
3. Select Maximum number of transients to find. If previous transients are stored in the memory, the maximum
available will be reduced appropriately. The information is available on the screen.
4. In the Threshold of Current Difference section of the dialog box, select the appropriate option in the drop-down
menu. Available options are the same as those in the Threshold of Voltage Difference drop-down menu.
The configuration settings can be saved to a file by clicking the Save to File button and selecting a file name. A previously
saved configuration can be loaded by clicking the Load From File button. A present configuration in the instrument (if
connected) can be read by clicking Re-Read From Instrument.