3.2.6.3 INRUSH CURRENT THRESHOLDS
You can also set Inrush current thresholds from the Capture Mode screen To do this, press the
This screen defines the Inrush current thresholds for each recognized current probe. The thresholds can be the same for
all channels or different for some or all of them. To select the Inrush current threshold, proceed as follows:
1. Select Start Threshold, highlighted in yellow, by pressing
to select the appropriate value.
2. Use the ▲ and ▼ buttons to raise or lower the threshold value, and the ◄ and ► buttons to move to a selected digit
in the value.
3. Press
to validate the threshold value.
4. Press ▼ to highlight the Hysteresis field. The selected field is highlighted in yellow. Press
the field.
5. Use the ▲ or ▼ button to raise or lower the value.
6. Press
to validate the hysteresis percentage.
To return to the Configuration screen, press
You can configure the current thresholds in mA, A, or kA.
Power Quality Analyzer PowerPad
Find Quality Products Online at:
.
®
III Model 8435
GlobalTestSupply
www.
Figure 3-16
. The navigation arrows appear in the field to enable you
.com
button.
to display the arrows in
sales@GlobalTestSupply.com
29