Measurement
Simple
RMS
Harmonic
Voltage
(order n ≥ 2)
Compound
SR clamp
MR clamp
MN93 clamp
SL261 clamp (10mV/A)
MN193 clamp (100A)
or 5A adapter
RMS
Harmonic
SL261 clamp \(100mV/A)
Current
(order n ≥ 2)
MN193 clamp (5A)
or 5A adapter
AmpFlex
AmpFlex
(1)
In 1000V
, CAT III, provided that the voltage between each of the terminals and earth does not exceed 1000V
rms
(2)
2-Phase (opposite phases) – same note as (1).
(3)
RMS value of the fundamental.
Power Quality Analyzer PowerPad
Find Quality Products Online at:
Measurement range
(with unity ratio)
Min
1V
1V
1A
0.2A
0.1A
0.01A
& MiniFlex
®
®
0.005A
& MiniFlex
10A
®
®
®
III Model 8435
GlobalTestSupply
www.
Resolution
(with unity ratio)
Max
0.1V
V < 1000V
1200V
(1)
1V
V ≥ 1000V
0.1V
U < 1000V
2400V
(2)
1V
U ≥ 1000V
0.1A
A < 1000A
1200A
1A
A ≥ 1000A
240A
0.1A
0.01A
A < 100A
120A
0.1A
A ≥ 100A
0.001A
A < 10A
12A
0.01A
A ≥ 10A
6A
0.001A
0.1A
A < 1000A
6500A
1A
A ≥ 1000A
.com
Accuracy
±(2.5 % + 1V)
±(2.5 % + 1V)
±(2% + (n x 0.2%) + 1A)
n ≤ 25
±(2% + (n x 0.5%) + 1A)
n > 25
±(2% + (n x 0.2%) + 1A)
n ≤ 25
±(2% + (n x 0.5%) + 1A)
n > 25
±(2% + (n x 0.2%) + 0.1A)
n ≤ 25
±(2% + (n x 0.5%) + 0.1A)
n > 25
±(2% + (n x 0.2%) + 0.01A)
n ≤ 25
±(2% + (n x 0.5%) + 0.01A)
n > 25
±(2% + (n x 0.2%) + 0.01A)
n ≤ 25
±(2% + (n x 0.5%) + 0.01A)
n > 25
±(2% + (n x 0.3%) + 1A + (Af
n ≤ 25
±(2% + (n x 0.6%) + 1A + (Af
n > 25
.
rms
sales@GlobalTestSupply.com
x 0.1%))
(3)
rms
x 0.1%))
(3)
rms
97