14.4 Disinfection

Your system should be cleaned and disinfected at least once a year to
eliminate any bacteria that are possibly in the system. We recommend
that you carry out cleaning and disinfection shortly before the time that
the filter cartridge must to be replaced.
Use cleaning solutions as follows:
MICRO-Chlor Granulate, 1 box, article no. 09.2202 (Europe only)
Cleaning Solution, 1 syringe, article no. CMX 25 (US-market only).
Please observe the information given in the safety data sheet supplied
with Micro-Chlor disinfectant to avoid possible health hazards!
Proceed as follows to disinfect your system:
1.
Disconnect the Smart2Pure from mains supply.
2.
Open the lid of the storage tank, pour the contents of a box respectively a syringe of
cleaning solution into the water-filled tank and then close the lid.
3.
Connect the system with mains supply and let it run for 1 hour in normal operation to
effective bacteria killing.
4.
Discard two tank fillings of water.
5.
Close the dispensing valve, disconnect the system from mains supply.
6.
Change the filter cartridge as described in the Operating Instructions supplied with the
system an if necessary the pre-treatment filter and the sterile filter.
7.
Connect the system with mains supply.
8.
Fill the tank completely once and discard the water produced from this tank filling.
Before dispensing water from the system, let water run out for approx.
15 minutes. The system is then ready for use.