Minimum Frequency
As necessary

Authorized service personnel

Minimum Frequency
12 months
2076152-001
Maintenance
Install new cylinder gaskets on cylinder yokes.
Replace the absorbent in the canister.
Empty and clean the overflow trap on the optional suction regulator.
Replace the circuit O2 cell. (Under typical use the cell meets
specifications for 2 years. )
Replace the autoclavable flow sensors. (Under typical use the sensor
meets specifications for a minimum of 1 year.)
Calibrate the airway modules every 6 months or when there are
indications of errors in the gas readings. Calibrate airway modules that
get extensive usage every 2 months.
Inspect and clean the fan filters (display unit, rear side of the machine
and airway module).
Empty D-fend Pro water trap on the airway module.
Drain the vaporizers and discard the agent as instructed in the
vaporizer User's Reference Manual (This is not necessary for the
Tec
6 series vaporizers.).
This is the minimum level of maintenance recommended. Local
regulations may contain additional maintenance requirements.
Comply with all local regulations which meet or exceed this minimum
level of maintenance.
.
Maintenance
Have an authorized service representative complete the scheduled service
maintenance checks, tests, calibrations, and parts replacements as defined
in the Technical Reference manual.