3.3 Door Size Requirements

Minimum door sizes also apply to hallways and elevators. For additional details, refer to
Shipping
Information.

3.3.1 Door Height

The minimum door height (to accommodate Innova positioner on its dolly) is 1.980 m (78 in). If
the height is limited to 2 m (79 in), you will need a fall over cabinet.

3.3.2 Door Width

The minimum door width needed (to accommodate the Innova LC shipping dolly) is:
1.165 m (46 in) with protective side rail,
1.096 m (43 in) with one protective side rail removed on site.
NOTE:
Door widths are based on a
corridor. Calculations need to be made for accommodation of equipment through
narrower corridors.
Chapter 1 General Requirements
Innova
IGS 520, Innova
TM
straight–in approach requiring a 2.44 m (96 in) wide
IGS 530, Innova