Auxiliary Output

Use the Aux Output dialog to configure the output of the Aux Out port.
Auxiliary Output
Trigger Enabled sends a pulse when the trigger is ready (Ready indicator lit), but not necessarily
triggered. It can be used as a gating function to trigger another instrument. Enter the desired pulse
Amplitude.
Trigger Out sends a pulse upon a trigger (TRIG'D Indicator lit). Enter the desired pulse Amplitude.
Pass/Fail generates a pulse when Pass/Fail testing is active and conditions are met. Enter the
desired Pulse Duration.
Fast Edge provides a fast edge signal that can be used to deskew multiple channels.
Off disables auxiliary input/output.
Calibration Output
Choose the signal to be output from the Cal Out hook on the front of the instrument:
Square wave signal. Enter the wave Frequency and Amplitude into 1 MΩ, or choose to Set to 1
l
kHz, 1 V Square Wave.
Reference DC Level. Enter an Amplitude into 1 MΩ.
l
Off disables calibration output.